J.U.M.'s FPM 605 reliably detects diesel oil,
jet fuel and all other PAH-containing oils in water. The FPM
605 is a continuous, on-line, process or field monitor which
provides the reliable measurement of low to medium ppm concentrations
of dissolved or emulsified oil in water. It is used for applications
which require oil monitoring in treated and untreated water streams.
This could be from industrial wastewater, groundwater, process
water, produced water and storm/rain water run-off.
Continuous monitoring of oils as PAH
(Poly- Aromatic Hydro- carbons) is a reliable,
effective, and well recognized method to improve the process
management for treating, discharging and detecting oils in water.
Compared to the laboratory grab sample analysis method, the technique
of continuous on-line monitoring provides cost effective, remote
and operator unattended measurement of poly aromatic hydrocarbons
in water.
The FPM 605 is based on the chemical principle
that poly aromatic hydrocarbons absorb light energy at one UV-
wavelength (called exication) and emit light energy at some longer
UV- wavelength (called emission). This process is named UV fluorescence.
Such fluorometer only detects hydrocarbons (and other materials)
which fluoresce with the selected pair of exication and emission
wavelengths. The fluorometeric detection of PAH's is a sensitive,
selective and specific method, which is nearly immune to air
bubbles and suspended solids that are often present in the water
to be measured. |
|
With our proprietary open cuvette flow cell,
the FPM 605 does not use a glass flow cell. The hydrocarbons
are detected in a stream of water which smoothly overflows our
flow cell. The water stream cannot contact, contaminate or foul
the optical windows which are used for the detection. No chemicals,
no pretreatment and no manual manipulation of the water sample
are required to monitor poly aromatic hydrocarbons.
Major Applications to be found in:
- Cooling Water
- Oil Coolers
- Oil Separators
- Condensers
- Heat Exchanger leak detection
- Fluorescin Tracer Detection in Cooling Water
|
|
Available Options |
|
AMU 65 |
Automatic range change |
|
CIM 65 |
RS 232 data output |
|
EAP 65 |
Air purge kit to meet
class I Division II (explosion profing) |
|
MWR 65 |
Built in mean value calculator |
|
PAS 65 |
Internal air supply with
pump and air purifier for air curtain |
|
RCA 65 |
0-20mA analog output instead
of 4-20mA |
|
RCC 65 |
Remote control range change |
|
RCI0 65 |
0-20 mA analog output,
galvanically isolated |
|
RCI4 65 |
4-20 mA analog output,
galvanically isolated |
|
|
|
Features & Benefits |
- Continuously Monitors the Dissolved and Emulsified
Oil in Water
|
- Quick Response: T90 < 12s
|
- Selectable Measuring Ranges 0-10, 0-100,
0-1000 ppm
|
- Analog Signal Out (opt. RS232)
|
- Optional Dual Alarm Threshold Indicator
|
|
|
|
|
- No Pretreatment or Chemicals Needed
|
|
|
|
|
|
|
Technical Data |
|
Method of operation |
UV fluorescence detector |
|
Dimensions (W x H x D) |
430 mm x 480 mm x 340 mm |
|
Weight |
approx. 18 kg (40 lbs) |
|
Detection range |
100ppb - > 1000 ppm (detection limits depend
on target analyte and matrix) |
|
Plumbing requirements |
Feed in: 2 - 5 liter / min, 2" unlimited
drain |
|
Response time |
< 12 seconds, depending on actual flow through
sample cell |
|
Operating temperature |
10 - 39°C, for extreme temperature conditions
contact manufacturer |
|
Ambient temperature |
10-39°C (50-102°F) |
|
Sample temperature |
Max 60°C |
|
Calibration |
Multi point (min. 2 point) |
|
Zero drift |
<2.0% FSD / 24h |
|
Span drift |
<2.0% FSD / 24h |
|
Linearity |
1,5% FSD |
|
Measuring ranges (ppm) |
either 0-100, 0-1.000, or 0-200, 0-2.000ppm |
|
Analog outputs |
0-10VDC und 4-20mA |
|
Display |
3 1/2 digit |
|
Alarms |
Two STDP relay contact, adjustable 100% FSD;
OPTIONAL |
|
Sample pump (optional) |
approx. 2.5 l/min capacity. |
|
Zero and span adjust |
Manual on front panel |
|
Maintenance |
Light Source once a year; routine cleaning of
all sample containing plumbing |
|
Air consumption for air curtain |
approx. 1,5 liter per minute (built it air supply
OPTIONAL) |
|
Temperature control |
for cold weather conditions, digital PID-type
controller, OPTIONAL |
|
Power requirements |
either 230VAC/50Hz, 150 W or 115VAC/60Hz,
150 W |
|
|
|
J.U.M. reserves the right, at any time and without
notice, to change specifications presented in this data sheet
and assumes no responsibility for the application or use of the
devices described herein. |
|