Burner Air Supply, Model TJ 38 & Model TJ 36
 

 

Description:

The J.U.M. Engineering Burner Air Supplies are highly efficient air purification instruments for the production of hydrocarbon-free air as required for hydrocarbon analyzers and gas chromatographs using the flame ionization detector (FID) principle.

It takes in ambient air through a dual-stage diaphragm pump assembly. The air then is filtered in a 4 micron particle filter and routed through a cooling coil and a water trap to collect any moisture from ambient air. The organic impurities (hydrocarbons) are then adsorbed in the dual-stage purifier. These purifiers are filled with a special type of activated carbon and have to be recharged only every 12-18 months.

The burner air pressure is controlled by two front panel pressure regulators. A pressure relief-valve prevents stalling of the pump motors after e.g. a short current failure.

Available as in 19" rack mount (TJ 36) or as a portable table top version (TJ 38).

 

 

1 Technical Data

 
 
Main pump Diaphragm pump w. Viton® diaphragm
Booster pump Diaphragm pump w. Viton® diaphragm
Output pressure range 1 0-1.6 bar (0-0.16 MPa, 0-23 psig)
Output pressure range 2 0-4.2 bar (0-0.42 MPa, 0-60 psig)
Particle filter replaceable 4 µm plate
Particle filter efficiency 99.9%
Hydrocarbon adsorption efficiency 99.999%
Pressure accuracy better than 0.5%
Power Requirements 230VAC/50Hz, 200W
(115VAC/60Hz, 200W)
Ambient Temperature 5-43°C (41-110°F)
Dimensions (W x D x H) 260 mm x 400 mm x 155 mm TJ 38
483 mm x 460 mm x 132 mm TJ 36
Weight 8 kg (18 lbs)


Viton® is a registered trademark of DuPont Dow Elastomers

 

 

 

 

 

J.U.M. reserves the right, at any time and without notice, to change specifications presented in this data sheet and assumes no responsibility for the application or use of the devices described herein.
 