Portable All Heated Sample Diluter Model 108
 

 

Description:

The J.U.M. Engineering heated sample diluter allows the use of HFID total hydrocarbon analyzers in concentration ranges above some 10.000 ppm. The portable Model 108 is ideal for use in combination with our portable HFID analyzer Model 3-200.

The use of our Model 108 eliminates the socalled "Hang-Up effects" in your FID analyzer. In measuring ranges above some 10.000 ppm HFID analyzers will become non-linear. To overcome this problem it makes sense to install a sample diluter. The J.U.M. Model 108 has a fixed dilution rate of 1:10. However, it can be adjusted to the sample flow rate of the FID via a precision regulator

Like all J.U.M. sample handling products it comes equipped with a heated, backpurgeable sample filter and a heated sample valve. An automatic backpurge option is available to extend calibration- and maintenance cycles.

 

 

Standard Specifications:

  • Sample Filter: All stainless steel
  • Filter pore size: 2µm
  • Sample valve: All stainless steel
  • Purge air valve: All brass
  • Calibration gas valve: All brass, or to be specified
  • Oven Temperature: Adjustable 366-483K (200-410°F / 93-210°C)
  • Temperature controller: Electronic PID-type
  • Power Requirements: 115V/60Hz AC, 800W; 230V/50Hz AC, 800W (others on request)
  • Ambient Temperature: 278-316K (41-110°F / 5-43°C)
  • Dimensions: Width x Depth x Height: 10" x 16" x 6" (260mm x 400mm x 155mm)
  • Weight: 12 lbs (5.5kg)
 

 

J.U.M. reserves the right, at any time and without notice, to change specifications presented in this data sheet and assumes no responsibility for the application or use of the devices described herein.
 