Description:
The J.U.M. Engineering HFID Model VE7 is very
rugged and versatile 19" rack mount heated total hydrocarbon
analyzer for high accuracy, sensitivity and stability.
The Model VE7 uses a hydrogen flame ionization
detector (FID) in a heated oven to prevent the loss of high molecular
weight hydrocarbons and to provide reliable performance in the
analysis of trace level of contaminants in high purity gases,
air and other gases.
The permanent heated sample filter is cleaned
by backpurging and has a replaceable stainless steel mesh filter
disc. A rear adaptor-system allows cold-spot free coupling of
a heated sample line inside the heated oven without the need
of special tools.
Download VE 7 product brochure as pdf-file (877kB).
(requires Adobe Acrobat Reader 4.0)
Major Applications:
- Stack gas hydrocarbon emissions monitoring
- Catalytic converter testing
- Raw exhaust vehicle emissions
- Detection of trace hydrocarbon levels in
purity gases
- Monitoring of hydrocarbon contaminants in
air and other gases
- Carbon adsorption regeneration control
- Measuring engine combustion efficiency
- LEL monitor of solvent-laden air
- Cryogenics/liquefaction
- Clean room applications
|
|
|
1
Technical Data |
|
|
|
Method of Analysis |
Flame Ionization Detector |
|
Sensitivity |
Max. 0.1 ppm CH4 |
|
Response Time |
0.2 seconds |
|
Zero Drift |
<1.0% full scale / 24h |
|
Span Drift |
<1.0% full scale / 24h |
|
Linearity |
within 1% of selected range |
|
Oxygen Synergizm |
< 1.2% |
|
Measuring Ranges (ppm) |
0-10,100, 1.000, 10.000, 100.000, others on request |
|
Signal Outputs |
0-10VDC and 4-20mA |
|
Display |
3 1/2 digit |
|
Sample Pump |
approx. 2.5 l/min capacity @ operating temp. |
|
Fuel consumption 100% H2 |
20 ml/min @ 1.5 bar (22
psig) |
|
Fuel consumption H2He 40/60-Mix |
90ml/min @ 1.5 bar (22
psig) |
|
Burner Air Consumption |
NONE, built in burner air supply |
|
Oven Temperature |
190°C (374°F) |
|
Temperature Control |
electronic PID-type controller |
|
Power Requirements |
230VAC/50Hz, 850 W or 115VAC/60Hz,
850 W |
|
Ambient Temperature |
5-43°C (41-110°F) |
|
Dimensions (W x D x H) |
483 mm x 460 mm x 221 mm |
|
Weight |
approx. 22 kg (46 lbs) |
|
|
|
|
1 Available Options |
|
|
|
|
AMU 7 |
Automatic range change |
|
APO 7 |
External automatic programmable backpurge system
for the sample filter |
|
AZM 7 |
Automatic flame ignition |
|
CIM 7 |
RS 232 data output |
|
DCC 7 |
Dual concentration threshold alarm, individaully
adjustable |
|
ENGA 7 |
6-digit display, 0-100.000 ppm |
|
FOAS 7 |
Flame out control with automatic fuel shut off
valve |
|
ICM 7 |
Built-In NMHC Cutter, measure either THC or Methane-Only
with one analyzer |
|
LTO 7 |
Special option
for measuring low trace hydrocarbon levels |
|
MBP 7 |
Integrated bypass pump |
|
MWR 7 |
Built in mean value calculator |
|
PDA 7 |
Sample pressure monitor with alarm |
|
RCA 7 |
0-20 mA analog output instead of 4-20 mA |
|
RCC 7 |
Remote control range change |
|
RCI0 7 |
0-20 mA analog output, galvanically isolated |
|
RCI4 7 |
4-20 mA analog output, galvanically isolated |
|
TPR 7 |
Built In temperature controller for JUM heated
sample line Model TJ 100 |
|
|
Click here
for a detailed overview and description of all available HFID
options |