We offer a wide range of options for all of our FID analyzers.
This enables us to customize analyzers for nearly every hydrocarbon
application. |
|
|
- AMU: automatic range
change for 5 measuring ranges; range increases if 95% full scale
of selected is exceeded, or range decreases if measured value
falls below 10% full scale of selected
|
- APO: automatic programmable
purge sequencer/timer;purge time and purge sequence programmable
from 1 second up to 10 hours; no filter probe or pre filter necessary;
internal sample filter, sample line and sample probe are purged
at the same time
|
- AZM: automatic ignition and reignition of FID-flame;
8-bit CMOS microcontroller with RISC architecture
|
- CIM : RS 232C data output
|
- DCC: concentration alarm with 2 thresholds, both adjustable
up to maximum range full scale, 2 potential free contacts; NOT
AVAILABLE ON MODEL 3-200
|
- ENGA: six digit engineering unit display 0 to 100,000ppm
(or other units) with integrated RS 232 data output
|
- ICM: Internal non-methane hydrocarbon cutter for measuring
either THC OR METHANE-ONLY with the same analyzer
|
- MBP: integrated bypass pump inside the heated chamber;
eliminates sample pressure fluctuations of up to 3 bar (44 psi)
and therefore their influence on the measurement results; NOT
AVAILABLE ON MODEL 3-200
|
- MWR: built in mean value calculator, 8-bit microcontroller,
2 MHz clock, 12 bit AD converter, real time clock with battery
buffer, data output via RS 232 C interface
|
- PDA: sample pressure monitoring as indicator for sample
pump failure; air pump failure; pressure regulator, clogged sample
filter, clogged sample line or sample probe
|
- RCC: remote control range change (dry contacts) with
range identification (dry contacts)
|
- RCI0: 0-20 mA output, galvanically isolated
|
- RCI4: 4-20 mA output, galvanically isolated
|
- TPR: built in temperature controller for heated sample
line Model TJ 100
|
|
|
|
|
|
|
|
|