Heated FID Analyzer for Total Hydrocarbon Measurement in Live Hot and Wet Steam, Model 606
 

 

Description:

This HFID analyzer is especially designed for the measurement of total hydrocarbons in live hot and wet steam.

It comes fully equipped to be connected to the steam source via a heated sample line. A 16 ft. (5 meter) heated sample line comes standard with the analyzer.

The hydrocarbon in steam analyzer Model 606 measures the concentration of total hydrocarbons in hot and wet steam. It is based on the flame ionization detection method. By applying a high polarization voltage between the two electrodes around the flame, which is created by burning pure hydrogen in hydrocarbon-free air, an electrostatic field is generated. During this process a very small but still measureable ion-current is produced. If a gaseous hydrocarbon-sample is introduced into the flame the ion-current changes proportionally to the number of hydrocarbon-molecules which are introduced in a unit of time.

A sample pressure regulator which does not come in contact with the sample itself creates a constant backpressure towards the sample capillary. This capillary provides a steady flow of sample to the detector. This method of constant pressure regulation of hot sample gases is being successfully used by J.U.M. Engineering for over 24 years. One part of the airflow is diverted after the pressure regulator and routed through an adsorption unit where the air is being purified to then be used as burner air for the detector. A compact flow control module with miniature needle valves and precision pressure regulators allows easy and reliable optimization of the detector. The internal manifold system with automatic backpurge of the sample filter and sample line in combination with the exclusively for that purpose designed filter/valve module allow long term measurements with very low drift effects.

 

 

1 Technical Data
 
Method of Analysis Flame Ionzsation Detector
Sensitivity Max. 0.1 ppm CH4
Response Time 0.2 seconds
Zero Drift <1.0% full scale / 24h
Span Drift <1.0% full scale / 24h
Linearity

within 1% of selected range

Oxygen Synergizm < 1.2%
Measuring Ranges (ppm) 0-10,100, 1.000, 10.000, 100.000, others on request
Signal Outputs 0-10VDC and 4-20mA
Display 3 1/2 digit
Sample Pump approx. 2.5 l/min capacity @ operating temp.
Fuel consumption 100% H2 20 ml/min @ 1.5 bar (22 psig)
Fuel consumption H2He 40/60-Mix 90ml/min @ 1.5 bar (22 psig)
Burner Air Consumption NONE, built in burner air supply
Oven Temperature 190°C (374°F)
Temperature Control electronic PID-type controller
Power Requirements 230VAC/50Hz, 850 W or 115VAC/60Hz, 850 W
Ambient Temperature 5-43°C (41-110°F)
Dimensions (W x D x H) 483 mm x 460 mm x 132 mm
Weight approx. 20 kg (45 lbs)
 

 

Built in mean value calculator

1 Available Options

   
AMU 60 Automatic range change
CIM 60 RS 232 data output
DCC 60 Dual concentration threshold alarm, individaully adjustable
ENGA 60 6-digit display, 0-100.000 ppm
FOAS 60 Flame out control with automatic fuel shut off valve
LTO 60 Special option for measuring low trace hydrocarbon levels
MWR 60 Built in mean value calculator
PDA 60 Sample pressure monitor with alarm
RCA 60 0-20 mA analog output instead of 4-20 mA
RCC 60 Remote control range change
RCI0 60 0-20 mA analog output, galvanically isolated
RCI4 60 4-20 mA analog output, galvanically isolated
 

 

 

Click here for a detailed overview and description of all available HFID options
 

 

J.U.M. reserves the right, at any time and without notice, to change specifications presented in this data sheet and assumes no responsibility for the application or use of the devices described herein.
 