Heated Total Hydrocarbon Analyzer, FID Model 5-100
 

 

Description:

The J.U.M. Engineering HFID Model 5-100 is a low-cost 19" rack mount heated total hydrocarbon analyzer for high accuracy, sensitivity and stability.

The Model 5-100 uses a hydrogen flame ionization detector (FID) in a heated oven to prevent the loss of high molecular weight hydrocarbons and to provide reliable performance in the analysis of trace level of contaminants in high purity gases, air and other gases. If equipped with option LTO 51 the analyzer is ideal for measuring trace hydrocarbons in high purity gases (5 ranges from 0-1ppm to 0-10.000ppm). It is also equipped with an automatic fuel shut of valve that interrupts fuel gas flow when the FID flame goes out.

Sample has to be introduced to the analyzer in prssurized form (max 1.5 bar, 22 psig).

Therefore the Model 5-100 is ideal for integration in CEM-Systems.


Major Applications:

  • Stack gas hydrocarbon emissions monitoring
  • Catalytic converter testing
  • Raw exhaust vehicle emissions
  • Detection of trace hydrocarbon levels in purity gases
  • Monitoring of hydrocarbon contaminants in air and other gases
  • Carbon adsorption regeneration control
  • Measuring engine combustion efficiency
  • LEL monitor of solvent-laden air
  • Cryogenics/liquefaction
  • Clean room applications
 

 

1 Technical Data
 
Method of Analysis Flame Ionzsation Detector
Sensitivity Max. 0.1 ppm CH4
Response Time 0.2 seconds
Zero Drift <1.0% full scale / 24h
Span Drift <1.0% full scale / 24h
Linearity

within 1% of selected range

Oxygen Synergizm < 1.2%
Measuring Ranges (ppm) 0-10,100, 1.000, 10.000, 100.000, others on request
Signal Outputs 0-10VDC and 4-20mA
Display 3 1/2 digit
Fuel consumption 100% H2 20 ml/min @ 1.5 bar (22 psig)
Fuel consumption H2He 40/60-Mix 90ml/min @ 1.5 bar (22 psig)
Burner Air Consumption max. 300 ml/min @ 1.5 bar (22 psig) depending on fuel gas
Oven Temperature 190°C (374°F)
Temperature Control electronic PID-type controller
Power Requirements 230VAC/50Hz, 850 W or 115VAC/60Hz, 850 W
Ambient Temperature 5-43°C (41-110°F)
Dimensions (W x D x H) 483 mm x 460 mm x 132 mm
Weight approx. 12 kg (28 lbs)
 

 

1 Available Options

   
AMU 51 Automatic range change
AZM 51 Automatic flame ignition
CIM 51 RS 232 data output
ENGA 51 6-digit display, 0-100.000 ppm
ICM 51 Built-In NMHC Cutter, measure either THC or Methane-Only with one analyzer
LTO 51 Special option for measuring low trace hydrocarbon levels
MWR 51 Built in mean value calculator
RCA 51 0-20 mA analog output instead of 4-20 mA
RCC 51 Remote control range change
RCI0 51 0-20 mA analog output, galvanically isolated
RCI4 51 4-20 mA analog output, galvanically isolated
 

 

 

 

Click here for a detailed overview and description of all available HFID options
 

 

J.U.M. reserves the right, at any time and without notice, to change specifications presented in this data sheet and assumes no responsibility for the application or use of the devices described herein.
 

 send literature

 back to analyzer overview

site map 

home