Wall Mount Heated Total Hydrocarbon Analyzer, FID Model 3-400
 

 

Description:

The J.U.M. Engineering HFID Model 3-400 is a compact wall mount heated total hydrocarbon analyzer for high accuracy, sensitivity and stability. It is housed inside a rugged and sealed NEMA 4 enclosure.

The Model 3-400 uses our proven "VentDownTM" hydrogen flame ionization detector (FID) in a heated oven to prevent the loss of high molecular weight hydrocarbons and to provide reliable performance in the analysis of trace level of contaminants in high purity gases, air and other gases. The "VentDownTM" design of the detector allows easy maintenance and helps preventing blockage of the detector exhaust.

All sample wetted components are integrated into the heated oven. The oven temperature is controlled by a microprocessor controlled temperature controller.

The permanent heated sample filter is cleaned by backpurging and has a replaceable stainless steel 2µm mesh filter disc. The burner air supply is built in. No extra bottles for combustion air are needed.


Major Applications:

  • Stack gas hydrocarbon emissions monitoring
  • Catalytic converter testing
  • EPA method 25A compliance monitoring of source hydrocarbons
  • Raw exhaust vehicle emissions
  • Solvent recovery monitor of carbon bed break through
  • Monitoring of hydrocarbon contaminants in air and other gases
  • Carbon adsorption regeneration control
  • Measuring engine combustion efficiency
  • LEL monitor of solvent-laden air
  • Cryogenics/liquefaction
  • Clean room applications
 

 

1 Technical Data
 
Method of Analysis Flame Ionzsation Detector
Sensitivity Max. 0.1 ppm CH4
Response Time 0.2 seconds
Zero Drift <1.0% full scale / 24h
Span Drift <1.0% full scale / 24h
Linearity

within 1% of selected range

Oxygen Synergizm < 1.2%
Measuring Ranges (ppm) 0-10,100, 1.000, 10.000, 100.000, others on request
Signal Outputs 0-10VDC and 4-20mA
Display 3 1/2 digit
Sample Pump approx. 2.5 l/min capacity @ operating temp.
Fuel consumption 100% H2 20 ml/min @ 1.5 bar (22 psig)
Fuel consumption H2He 40/60-Mix 90ml/min @ 1.5 bar (22 psig)
Burner Air Consumption NONE, built in burner air supply
Oven Temperature 190°C (374°F)
Temperature Control electronic PID-type controller
Power Requirements 230VAC/50Hz, 850 W or 115VAC/60Hz, 850 W
Ambient Temperature 5-43°C (41-110°F)
Dimensions (W x D x H) 600 mm x 200 mm x 600 mm
Weight approx. 30 kg (65 lbs)
 

 

1 Available Options

   
AMU 34 Automatic range change
APO 34 Internal automatic programmable backpurge system for the sample filter
AZM 34 Automatic flame ignition
CIM 34 RS 232 data output
ENGA 34 6-digit display, 0-100.000 ppm
ICM 34 Built-In NMHC Cutter, measure either THC or Methane-Only with one analyzer
LTO 34 Special option for measuring low trace hydrocarbon levels
MBP 34 Integrated bypass pump
MWR 34 Built in mean value calculator
PDA 34 Sample pressure monitor with alarm
RCA 34 0-20 mA analog output instead of 4-20 mA
RCC 34 Remote control range change
RCI0 34 0-20 mA analog output, galvanically isolated
RCI4 34 4-20 mA analog output, galvanically isolated
 

 

 

Click here for a detailed overview and description of all available HFID options
 

 

J.U.M. reserves the right, at any time and without notice, to change specifications presented in this data sheet and assumes no responsibility for the application or use of the devices described herein.
 

 send literature

 back to analyzer overview

site map 

home