Description:
The J.U.M. Engineering HFID Model 3-400 is
a compact wall mount heated total hydrocarbon analyzer for high
accuracy, sensitivity and stability. It is housed inside a rugged
and sealed NEMA 4 enclosure.
The Model 3-400 uses our proven "VentDownTM" hydrogen flame ionization
detector (FID) in a heated oven to prevent the loss of high molecular
weight hydrocarbons and to provide reliable performance in the
analysis of trace level of contaminants in high purity gases,
air and other gases. The "VentDownTM" design of the detector allows easy maintenance
and helps preventing blockage of the detector exhaust.
All sample wetted components are integrated
into the heated oven. The oven temperature is controlled by a
microprocessor controlled temperature controller.
The permanent heated sample filter is cleaned
by backpurging and has a replaceable stainless steel 2µm
mesh filter disc. The burner air supply is built in. No extra
bottles for combustion air are needed.
Major Applications:
- Stack gas hydrocarbon emissions monitoring
- Catalytic converter testing
- EPA method 25A compliance monitoring of source
hydrocarbons
- Raw exhaust vehicle emissions
- Solvent recovery monitor of carbon bed break
through
- Monitoring of hydrocarbon contaminants in
air and other gases
- Carbon adsorption regeneration control
- Measuring engine combustion efficiency
- LEL monitor of solvent-laden air
- Cryogenics/liquefaction
- Clean room applications
|
|
|
1
Technical Data |
|
|
|
Method of Analysis |
Flame Ionzsation Detector |
|
Sensitivity |
Max. 0.1 ppm CH4 |
|
Response Time |
0.2 seconds |
|
Zero Drift |
<1.0% full scale / 24h |
|
Span Drift |
<1.0% full scale / 24h |
|
Linearity |
within 1% of selected range |
|
Oxygen Synergizm |
< 1.2% |
|
Measuring Ranges (ppm) |
0-10,100, 1.000, 10.000, 100.000, others on request |
|
Signal Outputs |
0-10VDC and 4-20mA |
|
Display |
3 1/2 digit |
|
Sample Pump |
approx. 2.5 l/min capacity @ operating temp. |
|
Fuel consumption 100% H2 |
20 ml/min @ 1.5 bar (22
psig) |
|
Fuel consumption H2He 40/60-Mix |
90ml/min @ 1.5 bar (22
psig) |
|
Burner Air Consumption |
NONE, built in burner air supply |
|
Oven Temperature |
190°C (374°F) |
|
Temperature Control |
electronic PID-type controller |
|
Power Requirements |
230VAC/50Hz, 850 W or 115VAC/60Hz,
850 W |
|
Ambient Temperature |
5-43°C (41-110°F) |
|
Dimensions (W x D x H) |
600 mm x 200 mm x 600 mm |
|
Weight |
approx. 30 kg (65 lbs) |
|
|
|
|
1 Available Options |
|
|
|
|
AMU 34 |
Automatic range change |
|
APO 34 |
Internal automatic programmable backpurge system
for the sample filter |
|
AZM 34 |
Automatic flame ignition |
|
CIM 34 |
RS 232 data output |
|
ENGA 34 |
6-digit display, 0-100.000 ppm |
|
ICM 34 |
Built-In NMHC Cutter, measure either THC or Methane-Only
with one analyzer |
|
LTO 34 |
Special option
for measuring low trace hydrocarbon levels |
|
MBP 34 |
Integrated bypass pump |
|
MWR 34 |
Built in mean value calculator |
|
PDA 34 |
Sample pressure monitor with alarm |
|
RCA 34 |
0-20 mA analog output instead of 4-20 mA |
|
RCC 34 |
Remote control range change |
|
RCI0 34 |
0-20 mA analog output, galvanically isolated |
|
RCI4 34 |
4-20 mA analog output, galvanically isolated |
|
|
|
|
Click here
for a detailed overview and description of all available HFID
options |