Description:
The J.U.M. Engineering HFID Model 3-300A is
a very compact 19" rack mount heated total hydrocarbon analyzer
for high accuracy, sensitivity and stability and uses a hydrogen
flame ionization detector (FID) in a heated oven, which is heated
to 190°C to prevent the loss of high molecular weight hydrocarbons
and to provide reliable performance in the analysis of trace
level of contaminants in high purity gases, air and other gases.
All sample wetted components are integrated
into the heated chamber.
The permanent heated sample filter is cleaned
by backpurging and additionally has a replaceable stainless steel
2µm mesh filter disc. The burner air supply is built in.
No extra bottles for burner air are needed.
Our special rear adaptor-plate system allows
cold-spot free coupling of a heated sample line inside the heated
oven without the need of special tools.
Download 3-300A product brochure as pdf-file (727kB).
(requires Adobe Acrobat Reader 4.0)
Major Applications:
- Stack gas hydrocarbon emissions monitoring
- Catalytic converter testing
- Raw exhaust vehicle emissions
- Detection of trace hydrocarbon levels in
purity gases
- Monitoring of hydrocarbon contaminants in
air and other gases
- Carbon adsorption regeneration control
- Measuring engine combustion efficiency
- LEL monitor of solvent-laden air
- Cryogenics/liquefaction
- Clean room applications
|
|
|
1
Technical Data |
|
|
|
Method of Analysis |
Flame Ionization Detector |
|
Sensitivity |
Max. 0.1 ppm CH4 |
|
Response Time |
0.2 seconds |
|
Zero Drift |
<1.0% full scale / 24h |
|
Span Drift |
<1.0% full scale / 24h |
|
Linearity |
within 1% of selected range |
|
Oxygen Synergizm |
< 1.2% |
|
Measuring Ranges (ppm) |
0-10,100, 1.000, 10.000, 100.000, others on request |
|
Signal Outputs |
0-10VDC and 4-20mA |
|
Display |
3 1/2 digit |
|
Sample Pump |
approx. 2.5 l/min capacity @ operating temp. |
|
Fuel consumption 100% H2 |
20 ml/min @ 1.5 bar (22
psig) |
|
Fuel consumption H2He 40/60-Mix |
90ml/min @ 1.5 bar (22
psig) |
|
Burner Air Consumption |
NONE, built in burner air supply |
|
Oven Temperature |
190°C (374°F) |
|
Temperature Control |
electronic PID-type controller |
|
Power Requirements |
230VAC/50Hz, 850 W or 115VAC/60Hz,
850 W |
|
Ambient Temperature |
5-43°C (41-110°F) |
|
Dimensions (W x D x H) |
483 mm x 460 mm x 132 mm |
|
Weight |
approx. 20 kg (45 lbs) |
|
|
|
|
1 Available Options |
|
|
|
|
AMU 33 |
Automatic range change |
|
APO 33 |
External automatic programmable backpurge system
for the sample filter |
|
AZM 33 |
Automatic flame ignition |
|
CIM 33 |
RS 232 data output |
|
ENGA 33 |
6-digit display, 0-100.000 ppm |
|
FOAS 33 |
Flame out control with automatic fuel shut off
valve |
|
ICM 33 |
Built-In NMHC Cutter, measure either THC or Methane-Only
with one analyzer |
|
LTO 33 |
Special option
for measuring low trace hydrocarbon levels |
|
MBP 33 |
Integrated bypass pump |
|
MWR 33 |
Built in mean value calculator |
|
PDA 33 |
Sample pressure monitor with alarm |
|
RCA 33 |
0-20 mA analog output instead of 4-20 mA |
|
RCC 33 |
Remote control range change |
|
RCI0 33 |
0-20 mA analog output, galvanically isolated |
|
RCI4 33 |
4-20 mA analog output, galvanically isolated |
|
|
Click here
for a detailed overview and description of all available HFID
options |