Portable Heated Total Hydrocarbon Analyzer, FID Model 3-200
 

 

Description:

The J.U.M. Engineering HFID Model 3-200 is a portable "all-in-one" heated total hydrocarbon analyzer for high accuracy, sensitivity and stability.

The Model 3-200 uses a hydrogen flame ionization detector (FID) in a heated oven to prevent the loss of high molecular weight hydrocarbons and to provide reliable performance in the analysis of trace level of contaminants in high purity gases, air and other gases.

The permanent heated sample filter is cleaned by backpurging and has a replaceable stainless steel mesh filter disc. It has a built-in burner air supply so there is no extra cylinder needed.

A special portable container for fuel- and spangas is also available. Europe only!

Download 3-200 product brochure as pdf-file (726kB).
(requires Adobe Acrobat Reader 4.0)
Get Acrobat Reader


Major Applications:

  • Stack gas hydrocarbon emissions monitoring
  • Catalytic converter testing
  • Raw exhaust vehicle emissions
  • Detection of trace hydrocarbon levels in purity gases
  • Monitoring of hydrocarbon contaminants in air and other gases
  • Carbon adsorption regeneration control
  • Measuring engine combustion efficiency
  • LEL monitor of solvent-laden air
  • Cryogenics/liquefaction
  • Clean room applications
 

 

1 Technical Data
 
Method of Analysis Flame Ionization Detector
Sensitivity Max. 0.1 ppm CH4
Response Time 0.2 seconds
Zero Drift <1.0% full scale / 24h
Span Drift <1.0% full scale / 24h
Linearity

within 1% of selected range

Oxygen Synergizm < 1.2%
Measuring Ranges (ppm) 0-10,100, 1.000, 10.000, 100.000, others on request
Signal Outputs 0-10VDC and 4-20mA
Display 3 1/2 digit
Sample Pump approx. 2.5 l/min capacity @ operating temp.
Fuel consumption 100% H2 20 ml/min @ 1.5 bar (22 psig)
Fuel consumption H2He 40/60-Mix 90ml/min @ 1.5 bar (22 psig)
Burner Air Consumption NONE, built in burner air supply
Oven Temperature 190°C (374°F)
Temperature Control electronic PID-type controller
Power Requirements 230VAC/50Hz, 850 W or 115VAC/60Hz, 850 W
Ambient Temperature 5-43°C (41-110°F)
Dimensions (W x D x H) 300 mm x 580 mm x 204 mm
Weight approx. 18 kg (40 lbs)
 

 

1 Available Options

   
AMU 32 Automatic range change
APO 32 External automatic programmable backpurge system for the sample filter
AZM 32 Automatic flame ignition
CIM 32 RS 232 data output
ENGA 32 6-digit display, 0-100.000 ppm
FOAS 32 Flame out control with automatic fuel shut off valve
ICM 32 Built-In NMHC Cutter, measure either THC or Methane-Only with one analyzer
LTO 32 Special option for measuring low trace hydrocarbon levels
MWR 32 Built in mean value calculator
PDA 32 Sample pressure monitor with alarm
RCA 32 0-20 mA analog output instead of 4-20 mA
RCC 32 Remote control range change
RCI0 32 0-20 mA analog output, galvanically isolated
RCI4 32 4-20 mA analog output, galvanically isolated
 

 

 

Click here for a detailed overview and description of all available HFID options
 

 