Non-Methane Hydrocarbon/Total Hydrocarbon Analyzer, Model 109A
 

 

Description:

The J.U.M. Engineering Model 109A Non-Methane Hydrocarbon (NMHC) analyzer is a compact 19" rack mount analysis system using the FID (Flame Ionization Detector) principle. Its modular design allows easy maintenance.

There is no additional sample handling equipment required. The sample pump, the sample filter and all pressure and flow regulating components are built in. All sample wetted parts are operating at a temperature of 463K (374°F / 190°C) or above. To avoid reading errors due to calibration gas pressure-fluctuations the analyzer has an integral calibration gas overflow system before the sample filter.

The Model 109A also features our dual detector/dual electrometer design which provides three continuous, simultaneous signals of:

  • Total Hydrocarbons
  • Methane/Only
  • Total Hydrocarbons less Methane (NMHC)

The permanent heated sample filter is cleaned by backpurging and has a replaceable stainless steel mesh filter disc. A rear adaptor-system allows "cold-spot free" coupling of a heated sample line inside the heated oven without the need of special tools.


Major Applications:

  • Stack gas emissions monitoring
  • Ambient air monitoring to low concentration levels
  • Raw automotive and diesel exhaust analysis
  • Thermal reactor and combustor emissions monitoring
 

 

1 Technical Data
 
Method of Analysis Flame Ionization Detector
Sensitivity Max. 0.1 ppm CH4
Response Time Less than 1 sec. THC-channel
Less than 40 sec. CH4-channel
Zero Drift <1.0% full scale / 24h
Span Drift <1.0% full scale / 24h
Linearity

within 1% of selected range

Oxygen Synergizm < 1.2%
Measuring Ranges (ppm) 0-10,100, 1.000, 10.000, 100.000, others on request
Signal Outputs 3 x 0-10VDC
Display 3 x 3 1/2 digit
Sample Pump approx. 2.5 l/min capacity @ operating temp.
Fuel consumption 100% H2 40 ml/min @ 1.5 bar (22 psig)
Fuel consumption H2He 40/60-Mix 180ml/min @ 1.5 bar (22 psig)
Burner Air Consumption NONE, built in burner air supply
Oven Temperature 190°C (374°F)
Temperature Control electronic PID-type controller
Power Requirements 230VAC/50Hz, 1250 W or 115VAC/60Hz, 1250 W
Ambient Temperature 5-43°C (41-110°F)
Dimensions (W x D x H) 483 mm x 460 mm x 221 mm
Weight approx. 25 kg (55 lbs)
 

 

1 Available Options

   
AMU 9 Automatic range change
APO 9 External automatic programmable backpurge system for the sample filter
AZM 9 Automatic flame ignition
CIM 9 RS 232 data output
ENGA 9 6-digit display, 0-100.000 ppm
FOAS 9 Flame out control with automatic fuel shut off valve
LTO 9 Special option for measuring low trace hydrocarbon levels
MBP 9 Integrated bypass pump
PDA 9 Sample pressure monitor with alarm
RCC 9 Remote control range change
 

 

 

Click here for a detailed overview and description of all available HFID options
 

 

J.U.M. reserves the right, at any time and without notice, to change specifications presented in this data sheet and assumes no responsibility for the application or use of the devices described herein.
 